Users Guide
450 SERIES
Digital Thermometers
Shop online at
MADE IN
TEMPERATURE INDICATOR
HOLD
ON
°F
1°
0.1°
OFF
ON
°C
305.4
°F
THERMOCOUPLE THERMOMETER
TYPE E
450 AET
®
Page view 0
1 2 3 4 5 6 ... 30 31

Summary of Contents

Page 1 - User’s Guide

www.omega.com e-mail: [email protected]’s Guide450 SERIESDigital ThermometersShop online atMADE INTEMPERATURE INDICATORHOLDON°F1°0.1°OFFON°C305.4°FT

Page 2 - Servicing North America:

7Section 2 Operation 2.1 Principles of OperationThe temperature indicator uses two ICs to digitizeand display the analog input signal.One IC perform

Page 3 - Servicing Europe:

82.2 PowerWhen battery voltage reaches about 6 volts, thedisplay will exhibit a flashing: “LOW BAT” indication.A minimum of 20 hours of operation time

Page 4

9the recesses or each corner of the case bottom.Step 2: Carefully pull the case halves straightapart so as not to damage the internal connectors.Step

Page 5 - Section 1 Description

102.3 OperationStep 1: Insert temperature sensor plug into theindicator input jack. Failure to insert a probe resultsin a overload condition of three

Page 6

11Step 5: Observe and or record the temperaturevalue.Step 6: Turn instrument off by pressing the OFFkey.

Page 7

12450 SERIESOFFHOLDONF.ONC.10.1FCF/CIndicationDisplayValueProbeInput JackLowBatteryVoltageIndicationTo HoldDisplayPress andhold key toretain lastreadi

Page 8

132.4 MaintenanceNormal maintenance of the indicator consists ofoccasional cleaning with a soft, damp cloth,replacement of the battery as required, an

Page 9 - 450 Range Table

14 Section 3 Calibration3.1 CalibrationUnits with serial numbers beginning with theletter N use the calibration procedure outlinedbelow. Older units

Page 10 - Section 2 Operation

15with a minimum accuracy of ±0.1 degreeFahrenheit or ±0.2 degree Centigrade.Step 1. Ensure the unit is turned off. Connect theoutput of the thermoco

Page 11

16NOTE: If you want to cancel the calibrationmode press OFF button in front of the unit.For T/C type K set simulator @ 1320 °CJ set simulator @ 960 °C

Page 12

Servicing North America:USA: One Omega Drive, Box 4047ISO 9001Stamford CT 06907-0047Certified Tel: (203) 359-1660 FAX: (203) 359-7700e-mail: info@omeg

Page 13

17calibration procedure, set simulator at .2°C insteadof 0 °C. Repeat the calibration procedure again.Step 6. Verify calibration once finished.

Page 14

1810251638947GNDRef J+IN-1.23V+3.7VRaRcRbRa - Full Scale Adj.Rb Zero Adj.Rc +Full Scale Adj.

Page 15 - Operation Features

19Calibration- Thermocouple Type K, J andTand EEquipment List• 4 ½ digit voltmeter• Voltage Calibrator:Resolution, 1uV; accuracy ±0.01% (±2uV)• Ice Ba

Page 16 - 2.4 Maintenance

20c. Remove the bottom case of the indicator.d. Adjust the voltage calibrator for an output of0.000mV and connect the minus lead from thecalibrator to

Page 17 - Section 3 Calibration

21polrity).j. The voltage measured at the “3.7V” test-point should be 3.007 times larger thanthe voltage measured at the “-1.23V” test-point. Adjust

Page 18 - the input

22Calibration- Thermistor TypeEquipment List• 4 ½ digit voltmeter• Precision decade box:Resolution, 0.01 ohm ; accuracy ±0.02%• Ice Bath, 0.00 °C ± 0.

Page 19

23at the “-1.23V” testpoint: disregardingpolarity indication, record the voltage.f. Move the DVM HI lead to the “3.7V”testpoint and record the voltage

Page 20

24Calibration- Platinum RTDEquipment List• 4 ½ digit voltmeter• Precision decade box:• Resolution, 0.01 ohm ; accuracy ±0.02%• Ice Bath, 0.00 °C ± 0.0

Page 21

25e. Move the DVM HI lead to the “3.7V”testpoint and record the voltage (disregardpolrity).f. The voltage measured at the “3.7V” test-point should be

Page 22

26Zero Calibrationa. Turn the indicator on by pressing the °Ckey.b. Set the indicator to the 0.1° range.c. Connect the ice bath probe to the indicator

Page 23

Benelux:Postbus 8034, 1180 LA Amstelveen The NetherlandsTel: +31 (0)20 3472121 FAX: +31 (0)20 6434643Toll Free in Benelux: 0800 0993344e-mail: sales@

Page 24

WARRANTY/DISCLAIMEROMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date o

Page 25 - Calibration- Thermistor Type

TEMPERATUREThermocouple, RTD & ThermistorProbes, Connectors, Panels &AssembliesWire: Thermocouple, RTD &ThermistorCalibrators &

Page 26

Table of ContentsSection 1: Description1.1 General Description……………21.2 Physical Description…………...31.3 Unpacking……………………… 31.4 Specification

Page 27 - Calibration- Platinum RTD

2Section 1 Description1.1 General DescriptionYour hand-held, battery operated digitaltemperature indicator is configured for compatibilitywith one of

Page 28

31.2 Physical DescriptionThe indicator measures 45 mm (1.8 inches) inHeight, 85 mm (3.3 inches) in width, and 145 mm(5.7 inches) in length. Weight is

Page 29 - Zero Calibration

4SpecificationsDisplay:Liquid crystal display, 7.6 mm (0.3 in.) digitHeightOperating temperature:0 to 50°C (32 to 122°F) 0 to 75% RelativeHumidityStor

Page 30 - WARRANTY/DISCLAIMER

5Maximum Common Mode Voltage:1000VAC(RMS), ±2000V PeakAccuracy at 22°C (72°F)1° range: ±0.8°C±0.1% rdg±1.5°F±0.1% rdg0.1° range: ±0.25°C±0.1% rdg±.45°

Page 31 - Shop online at www.omega.com

6450 Range TableTypeKTypeJTypeEType T10.110.110.1-189 C to +1372 C-308 F to +2501 F-139.8 C to +203.7 C-219.7 F to +398.7 F-194 C to +1200 C-31