6-8 Calibration NetScan Users Manual
Gain Calibration of Main Unit
Required Equipment*
-
Volts DC Calibrator
-
22-2 Twisted pairs (to connect calibrator to calibration card)
-
Digital Voltmeter, or Digital Multimeter
(used for verification of calibration voltages)
-
Calibration Card (installed in bottom slot, slot #1, of main unit)
*
see required specifications on page 6-2
&$87,21
Offset calibration of main unit must be completed prior to calibrating the main unit for
gain.
&$87,21
Use approved ESD precautions, including static-free work area and grounded wrist strap,
when handling circuit boards and electronic components. Failure to do so could cause
equipment damage due to electrostatic discharge.